Previous Page  9 / 16 Next Page
Information
Show Menu
Previous Page 9 / 16 Next Page
Page Background

MAXX

Mess- und Probenahmetechnik GmbH

11

Technical requirements

>

Sampling with gentle sample extraction and hermetically sealed sample

storage in order to identify solvent residues (volatile substances) in water

>

Sampling by means of a peristaltic pump

>

Only inert materials are used (glass, PTFE (Teflon), PVDF)

>

Thanks to fast plug-in couplings the sample containers can be easily

removed for further analysis

Technical requirements

>

Sampler with self-emptying function and 7 x 5 L bottles

>

The sampler is installed in a chlorine plant and thus only special materials

like PTFE (Teflon) PVDF and silicone can be used

>

The device has two separate GRP cabinets. The top part as well as the

distributor drive were supplied with compressed air so that the components

are protected against an aggressive atmosphere

Technical requirements

>

Sampling device with 6 x 20 L bottles, XY-distributor

>

Glass metering vessel, housing made of V4A stainless steel, coated

refrigerating machine and sample bottles placed on a telescopic drawer

>

To monitor and record the interior temperature there is an additional

temperature measurement inside the bottle compartment

Technical requirements

>

Sampler with self-emptying function, rinsing and 7 x 5 L bottles

>

Samples have to be extracted from a circular pipeline and under exclusion of air

>

Permitted materials are only PTFE (Teflon) and PVDF

>

Sample dosing is effected per bottle by means of a little metering vessel

with overflow

>

Via a valve the sample medium flows from the circular pipeline into one of the

metering vessels until the water level reaches the overflow and thus defines the

sample volume. The sample is drained into the assigned bottle by means of a

further valve.